Tiger  XT VOC Detector

Tiger  XT VOC Detector

Description

The Tiger VOC detector is a revolutionary handheld gas detection instrument for the rapid, accurate detection of volatile organic compounds (VOCs) within the harshest of environments.

Features
Improved Robust Design
Large Internal Gas Table
IP65 Ingress Protection
Fast Response Time & Clear Down

Specifications

 

Minimum Resolution*

1 ppb or 0.001 mg/m3

Maximum Reading**

20,000 ppm or 20,000 mg/m3

Response Time

T90 < 2 seconds

Accuracy*

± 5% display reading ± one digit

Linearity*

± 5% display reading ± one digit

Intrinsically Safe Approvals

II1GExiaIICT4Ga
Tamb = – 15 0C =Ta =+45 0C (with alkaline battery pack) ITS09ATEX26890X IECEx ITS 10.0036X
3193491 conforms to UL Std. 913, 61010-1 &
Certified to CAN/CSA Std. C22.2 No. 61010-1

Battery Life

Li-ion: life up to 24 hours continual use
Alkaline: (Duracell Procell MN 1500) life up to 8.5 hours continual use

Lamps

10.6 eV Krypton PID lamp (standard) and 11.7eV lamps available

Data Logging*

Including date / time: 120,000 points

Communication

Direct USB 1.1 connection

Calibration

Standard calibration 100 ppm isobutylene Custom calibration capability

Alarm

Flashing LED and 95 dBA at 300mm (12”) audible sounder Selectable vibrating alarm
Pre-programmed TWA and STEL*
Pre-programmed with over 480 gases*

Flow Rate

220 ml/min in ambient conditions (with blocked flow alarm)

Temperature

Operating: -20 to 60 0C, -4 to 140 0F (non Intrinsically Safe)
Humidity: 0-99% RH (non condensing)

Protection

Designed to IP65 (heavy rain)
C E 1180
EMC tested to EN61326-1:2006, EN50270:2006 & CFR 47:2008 Class A

Weight & Dimensions

Instrument without probe Width: 340 x Height: 90 x Depth: 60 mm (13.4 x 3.6 x 2.4”) Standard case 420 x 320 x 97 mm (16.5 x 12.5 x 3.8”) Instrument weight 0.72 kg (1.56lb) Packed weight: 5 .5kg (12lb)

Standard case

420x320x97mm(16.5×12.5×3.8”) Instrument weight 0.72 kg (1.56lb) Packed weight: 5 .5kg (12lb)

 

 

*Model and gas dependent.
**Maximum reading is achieved with certain analytes such as ethanol.
***Specifications are based on isobutylene calibrations at 20 °C and 1000mBar.
All specifications quoted are at calibration point and under the same ambient conditions.
**** Based on continuous running.
#For indicative measurement only. Quoted accuracy achievable up to 2,000 ppm. For more accurate detection, calibration around concentration of interest is recommended.
Tiger XT V1.0. This publication is not intended to form the basis of a contract and specifications can change without notice.

(Fence Electrode Technology, United States Patent, Patent No.: US 7.046,012 B2)