Hygrometer Calibration Systems

Hygrometer Calibration Systems

Description

Calibration Parameter: Dew point, Frost point, RH%, PPM

Features
Calibration Range: Dew point(-90—+20℃)
Frost point(-95—0℃)
RH(10 to 98% @ 0-70℃)
RH(10 to 98% @ 0-70℃)

 

 

Calibration Parameter: Dew point, Frost point, RH%, PPM

 

 

Traceable to: NIST / SI

 

 

Calibration Range: Dew point-90+20℃)

Frost point-950℃)

RH10 to 98% @ 0-70℃)

PPM0.5 to 100,000 ppmV

 

 

Dew Point Hygrometer

  • Dual mirror temperature sensors
  • Removable mirror module for calibration in liquid
  • Heated measuring heads and sampling systems
  • High pressures up to 20000 kPa
  • Temperature controlled sampling system
  • Fast response
  • Touch screen full color LCD user interface

 

 

 

Specifications

Measuring Ranges

Frost/Dew Point: -95...20 °C

Temperature: -50...100 °C

Sample pressure: 500...2’000 hPa abs.

 

Accuracy

Frost/Dew point (for calibrated range): ≤±0.1°C

Temperature: ≤±0.07°C

 

Reproducibility

Frost/Dew point: ≤±0.05°C

Temperature: ≤±0.04°C

 

Standard Features

Digital I/O: RS-232

Mirror cooling: 3-stage Peltier thermoelectric

Mirror temperature sensors: Dual Platinum Resistance Thermometer PRT (Pt-100)

Analog outputs: Two analog outputs, user programmable, -10...+10 V and 4...20 mA

Gas connections: VCR Cajon 1/4’’

Sample gas flow rate: 0...2 l/min

Sample gas circuit: Stainless steel, 316 / 316L, electro polished

Cooling: Refrigeration

ORIS: Yes

Internal pump: Optional

 

 

Generator

 

  • Traceable to NIST
  • Based on NIST Proven Two-Pressure Principle
  • Push Button (Keypad) Control
  • Automated Control of User Setpoints
  • ControLog® Automation Software
  • HumiCalc® with Uncertainty Software
  • Automatically Applies Enhancement Factors
  • Computerized Internal Transducer Calibration
  • Timed/Formatted Output to Printer
  • Battery Backed-up Real Time Clock
  • Backlit Liquid Crystal Display

 

 

 

Specifications

Frost / Dew Point Range: –95 to +10 °C

Frost / Dew Point Uncertainty: (–95 to –90 °C) 0.9 °C

(–90 to –80 °C) 0.5 °C

(–80 to –70 °C) 0.2 °C

(–70 to 10 °C) 0.1 °C

Parts Per Million Range: 0.05 to 12000 PPMv

Relative Humidity Range: 0.0002 to 50%

Saturation Pressure Range:Ambient to 300 psiA

Saturation Pressure Uncertainty (10-50 psiA): 0.05

Saturation Pressure Uncertainty (50-300 psiA): 0.30

Saturation Temperature Range: –80 to +15 °C

Saturation Temperature Uncertainty: 0.08 °C

Saturation Temperature Heating/Cooling Rate: 2 Minutes Per °C Average

Gas Flow Rate Range: 0.1 to 5 L/min

Gas Flow Rate Resolution: 0.02 L/min

Gas Type: Air or Nitrogen

Gas Pressure Rating (MAWP): 350 psiG

Test Port: 1/4 Inch Swagelok® Tube Fitting (6.35mm)

Gas Supply (External): 350 psiG, 5 L/min

 

 

Applications

Used by NIMI, EPRI and academic institute for calibrating hygrometer

Used by dew point instrument manufacturers, humidity sensor and vapor content meter manufacturers for calibrating